Specifications for Models 810B Spectrum Analyzers

Return to 810B Product Description

Frequency

 
Frequency Range 200 KHz -810Mhz(1800MHz , 1800C)
Span 20 Khz to 1 Ghz
Center Frequency 0Hz to 810MHz
Marker Error +/- 1% Span

+/- 280 x Freq Ref Err

R.B.W. 300KHz, 10KHz,4KHz

RBW Accuracy +/- 10 %
RBW Selectivity 60dB / 3dB 10:1, 4 Pole Filter
Drift <50 Hz / Min., 30 Min warmup
Reference Oscillator Freq Ref Error: +/- 200 Hz @4 Mhz

Video Bandwidths Fixed at each RBW

Amplitude

 
Measurment Range -120 dBm to -30 dBm
Max. Safe Input Power (CW) +0 dBm( Internal Attenuator Limit)
Display Calibration 2, 5, 10 and 15 dB / div
Display & Readout Power Accuracy Avg Err. +/-3.0 dB, +/-4.0 dB Max.

300 KHz RBW, 1 MHz to 1400MHz

Incremental Resolution +/- 1 dB
Frequency Response Flatness +/- 2 dB from 1MHz to 800MHz

3 to 6 dB roll off @ 810 MHz

Internal Attenuator Range 20 dB in two steps, 10/20
Spurious Free Dynamic Range 60 dB @4 KHz RBW
Residual Responses < -90 dBm with input terminated, 0 dB attenuation.
3rd Order Intercept -10 dBm (Preamp ON)

Time Related Specs

 
1st LO Settling Time < 2000 µS
Sweep Time Min: 200 msec . Max: > 5 min.

486 Based System @ 33 MHz

Video Digitizing Rate/ Resolution > 10Ksps, 8 Bits

External Input Characteristics

 
RF Input 50 Ohm, Type F, female
RF Input VSWR < 1.5:1 , 0 dB attenuation

Other Specifications

 
Power Requirement

.1 amp @ 12 volts, 1 amp @5 Volt

Weight <1 lbs(no internal battery)
Size(WxHxD, Inches) 3.75 x 1.00 x 13.75

System Requirements IBM PC or compatible with VGA monitor, 1.4 MbB floppy drive, 2 MB Ram , One Parrallel Port and DOS 4.0 or higher. A 386 processor or better is recommended.